Specifications | 60V output series | 100V Output Series | |||||
Model | IPS-DH10-6012 | IPS-DH10-6025 | IPS-DH10-6030 | IPS-DH10-6040 | IPS-DH10-6050 | IPS-DH10-1007 | IPS-DH10-10010 |
AC input | Single-phase 220V±10% | ||||||
Frequency 50Hz/60Hz | |||||||
Output voltage | 0-60V | 0-100V | |||||
Output current | 0-12.5A | 0-25A | 0-30A | 0-40A | 0-50A | 0-7.5A | 0-10A |
Output power | 750W | 1500W | 1800W | 2400W | 3000W | 750W | 1000W |
Power stability | ≦ 0.1%+10mV | ||||||
Load stability | ≦ 0.1%+30mV | ||||||
Voltage regulation accuracy | ≦ ±0.1%FS | ||||||
Steady flow accuracy | ≦ ±0.1%FS | ||||||
Ripple (rms) | ≤ 0.1%FS+10mV | ||||||
Voltage setting accuracy | ±0.1%FS+10mV | ||||||
Current setting accuracy | ±0.1%FS+10mA | ||||||
Voltage readback accuracy resolution | 5-digit meter. Accuracy: ±0.1%FS + 3 digits (10%-100% of rated value). | ||||||
1mV | 10mV | ||||||
Current readback accuracy resolution | 5-digit meter. Accuracy: ±0.1%FS + 3 digits (10%-100% of rated value). | ||||||
1mA | 0.1mA | 1mA | |||||
Display method | Broken code screen display | ||||||
Output overvoltage protection | The menu settings include O.V.P. protection; the protection value can be set arbitrarily, and the output will be turned off after protection is applied. | ||||||
Output overcurrent protection | The menu settings include O.I.P. protection; the protection value can be set arbitrarily, and output will be turned off after protection is enabled. | ||||||
Over-temperature protection | Built-in OTP protection, with a protection value of 85℃ ±5% (radiator temperature). | ||||||
Voltage setting | Menu settings, encoder knob input | ||||||
Current setting | Menu settings, encoder knob input | ||||||
Communication methods | RS-485/RS-232 (supports Modbus and SCPI protocols) | ||||||
Function | Constant voltage/constant current, foldback current protection, memory mode, list sequence loop, configurable ON/OFF time, pulse width output, 9999 cycles, three preset voltage and current sets (full width model), and key lock. | ||||||
Output polarity | The positive (+) and negative (-) outputs can be grounded arbitrarily. | ||||||
Heat dissipation method | Forced air cooling | ||||||
Operating environment | Designed for indoor use, with a temperature range of 0℃ to 40℃ and a humidity range of 10% to 85% RH. | ||||||
Storage environment | Temperature: -20℃~ 70℃ ; Humidity: 10% ~ 90% RH | ||||||
Size and weight | W215*H44*L450 Approximately 2.8kg | W430*H44*L450 Approximately 3.5kg | W215*H44*L450 Approximately 2.8kg | ||||
Note: All values are typical and will be exceeded under different loads or other conditions. Accuracy is measured under load-sensing conditions; actual values may deviate from rated values due to temperature errors and component aging. | |||||||
Note: FS represents full scale. The three digits indicate that the last digit of the display will jump three times. | |||||||